NSF Soybean Functional Genomics
Vodkin Laboratory, University of Illinois
University of Illinois at Urbana-Champaign logo
NSF Home Overview Investigations

Workshop
Protocols

Soybean
EST Project

NSF Soybean Functional Genomics Project
Vodkin Laboratory, University of Illinois

Laser Scanning, page 1-3
Robin Shealy

 

Laser Scanning

Overview

The role of imaging systems in microarray analysis is to gather data about the level of hybridization of labelled probes to microarray spots. There are two main types of imaging systems: charge-coupled devices (CCD) and laser scanning systems. Laser scanning systems will be considered in this summary.

Laser scanners have one or more (usually two) lasers, most often of the HeNe variety (Helium/Neon), tuned to specific wavelengths determined by the dyes that are to be detected. Cy3 and Cy5 are the most common dyes used in probe labelling; the lasers for a Scanarray 3000 system (GSI Lumonics, Inc.) emit at 543nm (green) and 633nm (red). These wavelengths are in the regions of highly efficient excitation of their target dyes and negligibly efficient emission of the dyes. The wavelengths with highest excitation frequencies (556nm and 650nm for Cy3 and Cy5, respectively) are not used because these are in a part of the spectrum where there is a high intensity of emission from the dyes. Consequently, the ambient light from the lasers would be interpreted by the signal processing component of the scanner as a signal from the fluorophores.

The laser beam width defines the resolution of an image; most have resolutions around 10µm. This small area is picked up by the detector, a photomultiplier tube (PMT) and converted into an image pixel. The pixel usually has a 16 bit bandwidth, giving a grayscale of 0 to 65,535 as possible values. Most laser systems are confocal; that is, they are set up to focus the brightest intensity of the beam at the surface of the slide. The advantages of a confocal beam are efficiency of fluorophore excitation and reduction of noise from fluorescing contamination above and below the surface of the slide. The disadvantages are a tendency to rapidly photobleach the dyes, especially Cy5, and a sensitivity to small departures from planarity in the slide surface.

Scanning of a slide is most often done for one dye at a time. Initially, a low power, low resolution (50µm) scan is done to tune the laser power and PMT collection efficiency. This is done to prevent saturation of signal (i.e., too many high intensity pixels have 65,535 as a value) and to equalize the overall signal between the two channels. After tuning these values, the high resolution scans are done, and a pair of image files (tagged image format [TIFF] or bitmap [BMP]) are stored for subsequent spot detection.

Qualitative considerations in fluorescence imaging are (1) pixel size relative to spot size, (2) sensitivity of the system to a given fluor concentration and detectivity of a given DNA concentration, and (3) signal-to-background (i.e., everything that is not a fluorescing spot).

Laser Scanning of a Microarray

This protocol describes how to scan a hybridized slide using a ScanArray 3000 scanner.

Starting the system. Shut down the scanning computer. Turn on the scanner, then start the computer. Open the Scanarray software. Two image windows will appear inside the Scanarray window, one for each channel.

Insertion of slide for reading. Click on the Eject Slide button. Insert the microarray slide array surface up into the slide carriage that appears

Specification of array extent. Select Start from the Acquire menu; the Acquire Image window will appear (refer to Figure 1). Set the scan size by selecting the Custom radio button. Enter 1.0mm as the X start position, 17.5mm as the Y start position to scan. In the X box, enter 20mm (the width of the array); in the Y box, enter 40mm (length of array).

Do the following for each channel. Complete both steps for channel 2 (the Cy5 channel), then repeat with channel 1 (Cy3). Always start with channel 2 in scanning.

Initial scan. To assess if laser power and PMT settings for the channel will yield overall intensities covering the range of intensity values (0-65,535), perform a Quick Scan (50 micron resolution). Set laser power to 65% of maximum and PMT voltage to 80% of maximum as an initial approximation. Check the Quick Scan box, select the channel being scanned, and click Acquire. Be sure that only a single channel box is checked; otherwise both channels will be scanned. The image from the channel will appear in one of the image windows as the scan is made.
Manual adjustment of scanner settings. Compare the overall intensities of the spots, and depending on what you see, make adjustments to the laser and PMT settings to compensate. For example, if the overall intensity is moderately low, increase laser power from 65% to 70% and PMT from 80% to 85%; if it is extremely low, increase laser power to 75%-80% and PMT to 90%-95%. Adjustments of this type will take some experience. Higher laser power will tend to bleach the sample faster and higher PMT will tend to cause background fluorescence to be picked up. Perform another Quick Scan or two to obtain good intensity, using this method; try not to do more than 2 or 3 overall for each channel. Cy5 bleaches faster than Cy3, and it has a lot of variability in how many times it can be scanned.

Adjustment of intensities between channels. Compare images after the above adjustments are made to see if overall intensities for the Cy3 and Cy5 images are about the same (do not look at individual spots to make this judgement, unless they are carefully calibrated control spots). If necessary, make an adjustment to the laser power and/or PMT for one of the channels to equalize intensity, and do a further Quick Scan.

High resolution scan. Select Start from the Acquire menu. Check to see if the scan start position and scan widths are as specified above. Set the laser power and PMT to the values found in the Manual Adjustment step above, set resolution to 10 microns, check both channel boxes, and uncheck the Quick Scan box. Select Acquire. Channel 2 will be scanned first, then channel 1. It takes about 10-15 minutes for each scan to be performed. Images will appear in the two image windows.

Finishing the scan. Save the two images as TIF or BMP files. TIF files are in greyscale, BMP files have artificial colors assigned according to a user defined palette. Quantitation of spots on the images is done using the TIF files.

 

* Department of Crop Sciences
* College of Agricultural, Consumer, and Environmental Sciences
* University of Illinois at Urbana-Champaign


Design by: Crop Sciences Computer and Web Support Group
Copyright © 2000 University of Illinois at Urbana-Champaign